GENERAL DESCRIPTION

FULLY and FULLY SMART
AUTOMATED CLINICAL CHEMISTRY ANALYSER
CODE RM4000 - RM4020

SYSTEM OVERVIEW

FULLY - FULLY SMART
AUTOMATED CLINICAL CHEMISTRY ANALYSER
CODE RM4000 - RM4020

Biochemical Systems International Srl reserves the right to change specifications without prior notice.

SOFTWARE

FULLY - FULLY SMART
AUTOMATED CLINICAL CHEMISTRY ANALYSER
CODE RM4000 - RM4020

SOFTWARE OVERVIEW
Discover the simplicity of operating with Windows© environment.
Skilled laboratory staff will be able to run FULLY - FULLY SMART within short time. System maintenance instructions are integrated in the software.

- External panel touch PC computer (only for FULLY)
- Easy selection menu by icons
- The position of reagents is automatically determined by the workplan withthe volumes required once the samples and reagents have been programmed
- 9 profiles of up to unlimited chemistries each to simplify your task of profile management
- Continuous working status monitoring
- Automatic flagging of abnormal test samples with programmable interpretative messages
- Walk Away capability on reagents and samples and autoresumnig on error options
- QC program with Levey-jennings plot
- Intelligent error management
- Integrated maintenance scheduling
- Multilanguage support
- Local Area Network integration facilities
- External laser printer

SOFTWARE IMPROVEMENT

Software Improvements for FULLY and FULLY SMART

New features allowed with version 2.00.
Release date: December 2014

- Full compatible with 64 bit operative systems

Software Improvements for FULLY VET

New features allowed with version 1.11A
Release date: 15th January 2007

- Enabled Walk Away/Skip capability for reagents, samples and wells found empty
- Enabled Autoresuming option on errors (NOTE: PID 6 firmware version should be 1.15 or more or 2.02 or more)
- Service: enabled button to move down the plunger to make syringe disassembling easier
- Improved status monitor


Biochemical Systems International Srl reserves the right to change specifications without prior notice.

TECHNICAL FEATURES

FULLY - FULLY SMART
AUTOMATED CLINICAL CHEMISTRY ANALYSER
CODE RM4000 - RM4020

MEASURING SYSTEM
Operating mode Absorbance, End-Point, Fixed Time, Kinetic, Multistandard
Throughput Up to 100 tests/hour
Light source Halogen lamp 12V 20W
Spectral Range 320-690 nm
Filter Wavelength 340, 405, 492, 505, 546, 578, 630 nm, one free position
Measuring Range From –0.200 to 2.500 O.D.
Flow Cell 18 µl
Incubation temperature 37°C ± 0.2°C
Minimum Reading volume 350 µl
Typical Reading volume 500 µl
SAMPLE HANDLING
Sample number 54 positions including samples, calibrators and controls/tray
Sample container Cup (1.2 ml)
Sample volume 2-200µl (1µl steps)
Automatic dilution Pre and post dilution ration: 1:2 – 1:20
REAGENT HANDLING
Reagent tray 20 bottles + 1 bottle for diluent
Reagent bottle 45 ml
Reagent volume Reagent 1 volume range: 30 to 1000 µl (1µl steps)
Reagent 2 volume range: 0 to 1000 µl (1µl steps)
Liquid detection Sensor
Reagent warming Pre-heated ARM
Reagent identification Position ID
Tray exchange Possible with reagent bottles
REACTION
Number of wells 144 wells (12 wells x 12 strips)
Well volume 1 ml
Well temperature 37 °C; ± 2 °C
DILUTER
Syringe 1000 µl, 1 µl steps
Accuracy ± 1% at 5 µl
EXTERNAL COMPUTER (only for FULLY)
Computer External panel touch PC
Computer description Intel ® Processor 2 GHz, 150 GB HD, 2 GB RAM
Operating system Microsoft Windows ®
Display Panel touch
Network adapter Built-in
Printer External laser printer
Output RS232
GENERAL
Power Supply AC115-230V 50-60Hz full range
Dimensions 720 (W) x 680 (D) x 530 (H) mm
Weight 55 kg

Biochemical Systems International Srl reserves the right to change specifications without prior notice.

ACCESSORIES

FULLY - FULLY SMART
AUTOMATED CLINICAL CHEMISTRY ANALYSER
CODE RM4000 - RM4020

Each FULLY - FULLY SMART analyser is shipped with a StartUp Kit composed by the following items


CODE DESCRIPTION QUANTITY
300206/S Sample Tray 2
300207/S Reagents Tray 2
591007 Reagents bottles 40
591010 Cup for reagent bottles 40
Labels for reagent bottles rolls 2
592004 Reaction Wells 45
592003 Sample Cups 500
592006 KODAK cups 20
591009/S1-S2 500ml bottles (Wash-Waste) 2
400094 Wasting tube (1 m) 1
Triton solution 15% 1:100 (500 ml) 1
129007 Needle cleaning kit 1
Enlarging tubes pin 1
080027 Spare fuses (5A, fast) 2
132031 Power cable 1
170020/C1 20W 12V Lamp 1
User manual 1
121417_2 Washing well for needles 1
USB key with Software and manual 1
180189 PC Keyboard 1
180190 Mouse 1
300205 Separated little part of the cover 1
Release protocol 1
Warranty sheet 1
Unpacking sheet 1

Biochemical Systems International Srl reserves the right to change specifications without prior notice.

HUMAN
BIOCHEMISTRY ANALYZER SEMI-AUTOMATIC
BIOCHEMISTRY ANALYZER AUTOMATIC
VETERINARY
BIOCHEMISTRY ANALYZER SEMI-AUTOMATIC
BIOCHEMISTRY ANALYZER AUTOMATIC